Validation according to EN ISO 10077-2:2003
The program package AnTherm fulfils all requirements raised in EN ISO
10077-2:2003
for the simulation program to qualify it as standard method for calculation
of heat flow through frames of windows, doors and shutters.
The EN ISO 10077-2:2003 "Thermal performance of windows, doors
and shutters - Calculation of thermal transmittance - Part 2:
Numerical method for frames" in its appendix D provides
the validation procedure by which a "thermal bridges
program" must conform to, to be qualified as sufficient for the
calculation of heat flow through frames conformant to the
standards. The application package AnTherm conforms in its
current version to all respective criteria and assessments
specified by EN ISO 10077-2:2003 and thus it can be used as
"standard method for calculation of heat flow through frames of
windows, doors and shutters"
Detailed validation and benchmark results are put collectively into
the calculation report
of validation examples from the EN ISO
10077-2:2003 with the application package Antherm.
-
Reference cases 1 through 7 (thermal conductance Lf2D
and thermal transmittance Uf of frame profiles)
By concerning uncertainties resulting from rounding there is full concordance
of results calculated by AnTherm with values set in EN ISO 10077-2:2003.
The differences between thermal conductance Lf2D
calculated by AnTherm to those specified in the
EN ISO 10077:2003 stay below 1.1%. The requirement set in EN ISO 10077-2, that
differences of calculated thermal conductance must not exceed 3.0%, is met by AnTherm
in full.
The differences between thermal transmittance Uf calculated by AnTherm
to those specified in the
EN ISO 10077:2003 stay below 1.7%. The requirement set in EN ISO 10077-2, that
differences of calculated thermal transmittance must not exceed 5.0%, is met
by AnTherm in full.
-
Reference cases 8 and 9 (thermal conductance Lf2D
and transmittance U of profiles)
By concerning uncertainties resulting from rounding there is full concordance
of results calculated by AnTherm with values set in EN ISO 10077-2:2003.
The differences between thermal conductance Lf2D
calculated by AnTherm to those specified in the
EN ISO 10077:2003 stay below 1.2%. The requirement set in EN ISO 10077-2, that
differences of calculated thermal conductance must not exceed 3.0%, is met by AnTherm
in full.
The differences between thermal transmittance U calculated by AnTherm to
those specified in the
EN ISO 10077:2003 stay below 0.5%. The requirement set in EN ISO 10077-2, that
differences of calculated thermal transmittance must not exceed 5.0%, is met
by AnTherm in full.
-
Reference case 10 (thermal conductance Lf2D
and linear thermal transmittance Ψ of glazing, spacer and frame
compound)
By concerning uncertainties resulting from rounding there is full concordance
of results calculated by AnTherm with values set in EN ISO 10077-2:2003.
The difference between the thermal conductance Lf2D
calculated by AnTherm to the value specified in the
EN ISO 10077:2003 stays below 0.2%. The requirement set in EN ISO 10077-2, that
differences of calculated thermal conductance must not exceed 3.0%, is met by AnTherm
in full.
The difference between the linear thermal transmittance Ψ calculated by AnTherm
to the value specified in the
EN ISO 10077:2003 stays below 2.4%. The requirement set in EN ISO 10077-2, that
difference of calculated linear thermal transmittance must not exceed 5.0%,
is met by AnTherm in full.
Read more here:
See also: Validation according to EN
ISO 10211:2007,
Main menu Help->Example Load...
|